Document
C92IP001EN-M_Brochure_MCRxx2series-compressé
Contenu du document
MCR
The
Modular
Compact
Rheometer
Series
MCR: Your future-proof rheometer.
Look forward to your
future applications …
Which accessories are
available to extend the
MCR’s functionality?
14!bbRD 15
16!bbRD 17
Which measuring systems
best suit your needs?
18 !bbRD 19
Specifications
10 !bbRD 11
What are your
advantages due to the
instrument’s design?
6 !bbRD 7
Which features are
available for easy, reliable
MCR operation?
4 !bbRD 5
What does
modularity mean?
8 !bbRD 9
How do the synchronous EC
motor’s capacities work for you?
12 !bbRD 13
Which temperature control
options do you require?
MCR rheometers offer
you one thing first and foremost:
An open range of possibilities.
Modular.MCR rheometry builds on your applications.
Whatever your rheological requirements are and will be in the future – MCR rheometers are efficiently and comfortably adapted to meet your needs. The intuitive application software and patented features
like Toolmaster™, an automatic tool recognition and
configuration system, make sure of this.
Changing a cone-plate for a concentric-cylinder measuring system is just as easy as integrating a new temperature device or extending your rheometer’s testing capabilities
with a wide range of application-specific accessories.
Compact. MCR rheometry builds on your working day.
The space-saving MCR rheometers are designed with a specific focus on ease-of-use, with all components incorporated into one simply installed unit that easily fits
on a standard laboratory table.
‘Compact’ use of your time is guaranteed: The TruGap™ system for automatic gap control, the T-Ready™ feature
for controlling the actual sample temperature, and of course the speed and precision of the dynamic EC motor itself – these and other MCR characteristics ensure
efficient rheological operation.
Rheometer. MCR rheometry builds on technological innovation.
Benefit from rheological highlights such as the air-bearing- supported, synchronous EC motor, the dynamic TruRate™
sample-adaptive motor controller, the normal force sensor integrated in the air bearing, TruStrain™ real-time position
control, continuously improved electronics and numerous
other features that ensure the MCR series’ world- renowned rheological peak performance.
Anton Paar’s Modular Compact Rheometer series: From routine quality control applications to high-end research
& development.
MCR: The Modular
Compact Rheometer Series
SmartPave 102 HTR
RheolabQC FRS
DSR
SmartPave 92
MCR 72
WESP
MCR 92
WSP
MCR 102 MCR 302
MCR 702
MCR 502 MCR 502 WESP
5
4
Features for
Comfort and Efficiency
Automatic recognition of measuring and environmental systems: Toolmaster™
The MCR’s modular concept builds on the simple exchange of measuring systems and environmental
systems. This exchange basically organizes itself in a
6G?ixX9v#P4G g)x#q!0p#v1T?PU?'7???`??2v 8? 1? automatically does the work for you, without any
selections in the software.
Toolmaster™ is the only completely automatic tool recognition and configuration system for rheology. It
recognizes measuring and environmental systems as soon as these are connected to the rheometer. Transponder
chips in each accessory’s control cable and in the measuring system contain all relevant data, such as
truncation, diameter, cone angle and serial number, and automatically transfer these to the application software.
Error-free documentation and perfect traceability
x e@v%u87hqf% A?frHs?'4?Rc@t? tDYD-
Easy fitting of measuring systems:
QuickConnect
QuickConnect additionally offers you mechanical ease-of-use: The quick-fitting coupling allows one-hand
connection of the measuring systems and ensures fast,
convenient system changes without the use of a screwing mechanism.
Intuitive control of your device:
MCR color display
With the MCR series’ color display you can manage the complete sample preparation procedure directly at the instrument. The softkeys under the screen offer you the
same functionality as a touchscreen, but without the risk
of damaging or contaminating the screen in harsh working environments. Physical properties such as normal force,
temperature and gap are displayed in a clearly arranged, accessible fashion. Permanent control of the measuring gap:
TruGap™
Errors in gap size due to thermal expansion or contraction directly influence the accuracy of results in parallel-plate and cone-plate measurements.
The TruGap™ system fully replaces these procedures: The gap is directly measured and precisely adjusted to the desired position independently of the temperature and
thermal expansion.
TruGap™ measuring systems are based on a magnetic induction principle. An AC current flows through the
primary coil in the lower plate, which induces a voltage in the secondary coil since the circuit is closed by an iron
disk in the upper measuring plate. Based on this voltage,
the gap size is consistently measured and adjusted.
Time-saving temperature certainty:
T-Ready™ Rheological measurements are strongly influenced by temperature. Therefore, in addition to accurate
temperature control, knowledge about the sample
temperature equilibration is essential. The new T-Ready™ feature employs TruGap™
functionality to precisely
determine when the desired sample temperature has been
reached. Unnecessarily extended waiting times before
tests are eliminated: T-Ready™ gives a green light so that
the test can be started as soon as the desired sample
temperature is reached.
New Paths for Your Applications
The RheoCompass Software
Your rheometer opens up a constantly growing number of measurement opportunities.
This calls for a navigation tool that gives you the complete overview as well as the exact
?bv?(E?8P%`? @S&0?rS ?S?f?r?`"W6?XH$YdX4ff??A?a"$1?Ei?qfw0(FP?W??0s?
up-to-date rheometer software available on the market.
Designed for intuitive use, RheoCompass enables application-oriented template filtering,
customized test and analysis definitions, highly simplified data retrieval and much more.
7
6
The Key to Accuracy
The EC Motor Technology
The air-bearing-supported synchronous ?V?SgW"bY8U@5 ?(&tYExBy?4WFW 1?Ei?6VTQ?component of the MCR rheometer series.
Whether you perform zero-shear viscosity determinations of low-viscosity polymer
solutions or measure highly viscous magnetorheological fluids at high shear rates
??0s?I@?1x?P'?V?)4hq` motor of the MCR series ensures accuracy
across a wide viscosity range – from solids to
liquids with viscosities lower than water.
The rotor of the EC motor drive is equipped with permanent magnets. In the stator, coils with
opposite polarity produce magnetic poles. The magnets in the rotor and the stator coils attract
each other, so that a rotating flux of current in the coil windings produces a frictionless
synchronous movement of the rotor.
The torque of the motor is set and measured via the input current to the stator coils. Due to its unique design the EC motor features a linear relation between the torque and the input
current to the stator coil, which is advantageous
for precise torque control and measurement.
These and other motor characteristics considerably benefit your rheological measurements.
The optical encoder
The high-resolution optical encoder based on data oversampling technology enables the measurement ??0s?$S?twA8?@cda?5!RAXwc20Bb? ?%$Q?8 4#pD0
Combined with TruStrain™ this provides the basis for comprehensive studies of samples with weak structures.
The speed of completely digital control
The use of the most recent processor technology in the B?( 3S?? %?38Gc?1?Ei?DSXcda?@FT6e26?R????0s?increases the efficiency of transient tests. Due to the
additional memory, important signals such as the torque
and deflection angle are recorded and processed with
higher signal density.
?S?@T??1?YY@%PPXh?8 ?0uG? &7?RI?6 7(X??X?A?x? 5#QR??6#?p)x#q!7(X??1?Ei?B?( 3S?? being the first rheometers ever based on this technology.
As a result of constant optimization, MCR rheometers are now also controlled with digital current sources. The low
fluctuation and high performance of these sources further increases the accuracy of MCR torque measurement and
control.
The air bearing
Two air bearings support the motor: A radial air bearing centers and stabilizes the shaft and the axial air bearing holds the weight of the rotating parts. This established air
bearing technology is independent of external influences and therefore applicable without further electronic control.
Continuously optimized for rigidity, drift stability and robustness, the MCR rheometers’ air bearing technology
together with improved torque scanning enables low-torque $ d0??8cPBb? ?%$Q?8 ?06D?Acda?%t?&G?sid
The normal force sensor
The high sensitivity and increased sampling rate of the normal force sensor integrated in the air bearing enables
normal force measurements during transient and steady-
state tests as well as static normal force measurements,
which are used for gap control and DMTA, tack or penetration tests.
The sensor employs an electric capacity method, precisely converting extremely small deflections in the air bearing into
the according normal force. Instead of enforcing additional
travel, the natural movement already present in the air bearing is used to measure the normal force.
The advantage of the sensor’s location in the air bearing: Normal force measurements are available for all
temperature devices and application-specific accessories.
Whatever your application, the normal force sensor is always immediately functional.
Motor characteristics Your rheological advantage
Instantaneous buildup of
magnetic field, no magnetic
induction Fast response times for
step rate and strain tests
No eddy current and heat
production in motor Permanent torque values up
%$Q7?G"(S?h
Linear relationship between
electromagnetic torque and
stator current, one single
motor constant Control and resolution of
lowest speeds, deflection
angles and torques,
TruStrain™
Known constant magnetic
field allows motor-controlled
blocking EbB?rimlock”: Motor-controlled
blocking during sample
trimming
9
8
Intelligent
DesignIntelligent
Control
Space to work: The housing
Ergonomics, functionality and durability were the central design goals of the compact MCR housing. All mechanical and electrical control components are incorporated into one single, simply installed unit that easily fits on a
standard laboratory table. You are given sufficient space
for sample loading, trimming and further preparation.
The mechanical self-alignment of all accessories ensures comfortable work with the modular MCR system. In addition, customized housing solutions are available,
e.g. for adaptation to a confocal microscope.
Maximum rigidity: The frame Y%W%$Q1?Ei?B?( 97x??dD y?a@#?Q?$??? environmental temperature cannot influence the results
cda?4??d 4?`x?P'H$YB?( 3S??D4?Rc@t? ts1r? 7(X???8 ?0 steel frame for optimized mechanical and thermal stability.
In addition to the low elastic compliance of the steel itself, the Isolign'BX? Piezo Flange actively compensates the
residual compliance.
IsoLign'BX?: Nano-scale precision
The unique IsoLign'BX? Piezo Flange enables nanometer- scale gap size changes as small as 10 nm by three
Piezo elements in the rheometer’s bottom flange. This system provides additional measuring gap constancy – an especially vital feature for long-term measurements
at low torques – and actively compensates the entire
system’s axial compliance in transient tests. In tests across
vast temperature ranges, the measuring gap is kept constant within lower tolerances than ever before.
Connections
x?P'r4dH Ccda?1?Ei?B?( 3S?? WFW !$yr%`? 6ev?? by the instrument’s interfaces. The result: More options for
flexible work.
#F??R USB interface for direct communication with computer
#F??R Ethernet interface for direct or network communication
#F??R% A? TQ?.vC?'DchA#?D#tP%6w??uSh
#F??R 2 auxiliary inputs for reading external devices
#F??R Thermocouple interface for temperature reading
#F??R Pt 100 interface for temperature reading
#F??R% A? Ap?9?w???(BIBA#?D$S?twd p%6w??b`?Y=
#F??R Connector for magnetic valve TruRate™
The MCR series’ TruRate™ sample-adaptive controller
intelligently adapts to the sample conditions at hand.
Without prior information on the sample, and without any
pre-testing, sample strains, shear rates or stresses are
precisely controlled. The desired settings are achieved
in minimum time; no additional software selections are
required.
TruRate™ swiftly adjusts to the desired shear rate step
or step in strain without overshoots – for accurate
investigations of all kinds of samples.
TruStrain™
Strain-controlled oscillatory tests with common stress-
$S?twx#(??6RW#gH?#%`? @S&0@t? t?8 ?0? /
game”, including several oscillation cycles and subsequent
adjustments, to reach the desired strain amplitude.
TruStrain™ takes a different approach: Instead of
amplitude control, it employs real-time position control
%Q!8e%PPXh1?Ei?9`xf$7p4???%%B6$UR20Td)49?Dr#xS
ensures more efficiency and drift-free measurements at
smallest torques and strains.
TruStrain™ adjusts to the desired strain directly on the
sine wave, and the measuring system directly follows
this required change in strain during each individual
oscillation cycle. This means you are able to preset and
control precisely sine-shaped strains both within the
?h?""AQr?&?)`a6 t (9?0p#v!$yr0p#v$???
6hF25e77?%Rd8?5?ic1???!?&`
close observation of intercycle processes, the application
software optionally displays oscillatory waveforms and
Lissajous diagrams.
TruStrain™ is especially valuable for oscillatory
measurements on complex fluids such as gels, emulsions,
suspensions, colloids, surfactant solutions, lubricating
grease and foams.
All rheological parameters
The electric motor torque, the forcing frequency (set
W2D??0s?1?Ei?7 ' A!pcda?P??8w y?a@&6Fdi??py?a@1?Ei?
deflection angle and the phase shift between the electric
SgW2u"?i?8??0s?1?Ei?B7'?%`?E6?5y()BFey
can be used to determine all rheological parameters.
11
10
Anton Paar offers a wide range of modular
temperature devices tailored to specific rheological
requirements – from -160 °C to 1000 °C, from
low-viscosity liquids to highly elastic solids, from
traditional rheological tests to DMTA measurements. All systems are easily exchanged within the MCR
series and ensure precise temperature control and
uniform temperature distribution for all applications.
Make your first selection here. Anton Paar’s temperature devices are based on the
physical principles of conduction, convection and
radiation.
Paste-like materials
Soft solids
Solids Solids
Melts
Powder/reactive systems
Modular Temperature Control …
$X 6GF't?FRhxH4?0?%$Q? ??Bp?
* Room temperature
Low-viscosity liquids
Viscoelastic liquids
Gel-like materials
13
12
6 `(!6r?P2 #?1T?Pfu8B` ?F?#W??#W??
6 `(!1?g1T?P&?T54ri ?1u4P`?Y B BRG?? TPATU?pt5 6 `(!6r?PwWp&?T54ri ? ?F?47d(?B BRGBBuU?47d(?
@RU"?2 #?1T?Pfu8B` ?F?#W??#W??
@RU"?2 #?1T?Pfu8B` ?F?#W??#W??B BRG?? TPATU?pt5 6 `(! 1T?m &?T54ri ? (? #W??
wWp&?T54ri ?#F??R47d(?6 `(!1T?Pfu8B`#F??RE??`?#w@1T?P&?T5t5 CTD 600 MDRCTD 1000
6r?PBBuU?#w@$iv& HF@p?0@X ?F?#W??#w@$iv6 `(! #w@$iv& HF@p?0@X1u4P`(x ?#w@$iv6 `(! t1T?P1T$T@p?0@X ?F?#W??#W??B BRGBBuU?#W??#w@$iv6 `(!
6r?P#F??R#w@$iv& HF@p?0@X ?F?#W??#w@$iv6 `(!
6r?PX?hHP$iv& HF@p?0@X ?F?#W??#w@$iv6 `(! 6r?P ?F?#w@$iv& HF@p?0@X1u4P`(x ?#w@$iv6 `(!
6r?P1u4P`E??`?#w@$iv& HF@p?0@X#F??R#W??#w@$iv6 `(!6r?P1u4P`E??`?#w@$iv& HF@p?0@X#F??RE??`?#w@$iv6 `(!6r?P1u4P`??#w@$iv& HF@p?0@X??#W??#w@$iv6 `(!6r?P1u4P`#W??#w@$iv& HF@p?0@X1u4P`#W??#W??#w@$iv6 `(!
TQ? Cylinder
TQ? Conduction TQ?
Cylinder
TQ? Conduction TQ?
C P/ PP
TQ? Conduction
TQ? Convection
TQ? Radiation TQ?
CP/PP/CC/DMTA solid and
extensional fixtures
TQ? Convection
TQ? Radiation TQ?
C P/ PP
TQ? Conduction
TQ? Convection
TQ? Radiation TQ?
CP/PP/CC/DMTA solid and
extensional fixtures
TQ? Convection
TQ? Radiation TQ?
CP/PP/CC/DMTA solid
and extensional fixtures
TQ? Convection
TQ? Radiation TQ?
CC/PP
TQ? Convection
TQ? Radiation
TQ? Fully incorporated, truly Peltier-
temperature-controlled system
TQ? High heating and cooling rates
TQ? No vertical temperature gradient in
sample due to optimized thermal
transfer system
TQ? Counter-cooling by air or fluid
circulator
TQ? Temperature control for pressure
!t "CGV5?U5?U"W6?X)??`6 `(!6r?P2 #?1T?Pfu8B` ?F?#W??#W??3a TQ?
Especially suited for
measurements of low-
viscosity samples at high temperatures
TQ? High heating rates
TQ? Ideal for use with pressure
cell TQ?
Fully incorporated, truly Peltier-
temperature- controlled system
TQ? Tr u G a p 'BX? support
TQ? T-Ready 'BX? feature
TQ? Sliding rail for easy access and
sample trimming
TQ? Evaporation Blocker: prevents loss
of volatile solvents
TQ? Temperature-isolated hood
(h rA@c`dApA6?p)6r?PXFr)rA@c%`? C?E?x) &?@?C2@?`X%YTf&rA@c(?x) @t? t!45?!%YTf&GV3a
TQ? Isolated hood according to
EN61010-1:2001 TQ?
Truly Peltier-temperature-
controlled convection oven
TQ? Tr u G a p 'BX? support
TQ? T-Ready 'BX? feature
TQ? DigitalEye CCD camera
function
TQ? Modular configuration
"%?BDMTA torsion/tension, Photo DMTA /UV, reaction kinetics, SER extensional
%`? )?(??1cwcpAG)?1i ?3a
TQ? Humidity Option
TQ? Temperature-isolated
jacket (hand-warm for
s rA@c(?x) GV% A?!45?!%YTf&'2?? TQ?
Ideal for measurements
of tablets, granules and powders
TQ? Sliding rail for easy access
and sample trimming
TQ? Temperature-isolated
hood according to
EN 61010-1:2001
)??`6?7PrA@c`dApA6?p)6r?PXFr)rA@c%`? C?E?x) &?@?C2@?`X%YTf&rA@c(?x) @t? t!45?!%YTf&GV3a
TQ? High heating rates
TQ? Cooling by gas, water or
liquid nitrogen TQ?
Modular configuration (DMTA torsion/tension, Photo
D M TA / U V, % A?%`?4GVrA@c!t "C@p?S??%% A??P'Y P9?d@p?Yxb?%YTf&0S& #% A?4p???0QX`t% A?GV4?Y@p?GV`dAp%YTf&S??%0eY`% A?s?A?'Y %YTf&!t "C&?@?C%YTf&i)w 1i ?3a
TQ? Tr u G a p 'BX? support
TQ? T-Ready 'BX? feature
TQ? Digital Eye CCD camera function
TQ? Pt 100 signal reflects true sample temperature
TQ? 1T?PGV0BG94h)GV%`?5rA@c@p?!45?!%`? GV6r?P'Y BYTf&&?@?C5?UrA@c@p?GV?Y?gC ?rA@c!t "C?PGV1R?04)??`6?7PrA@c`dApA6?p)6r?PXFr)rA@c%`? C?E?x) &?@?C2@?`X%YTf&rA@c(?x) @t? t!45?!%YTf&GV3a
TQ? Actively cooled jacket according to EN 61010-1:2001
TQ? Evaporation unit actively controls continuous flow of
liquid nitrogen: most stable temperature signal for
l ow-temperature applications
TQ? Gas-Chiller Option for cooling without liquid nitrogen TQ?
Most suitable for
measurements of glass
and metal melts
TQ? Thermocouple signal
reflects true sample
temperature
TQ? Temperature-isolated
jacket (hand-warm for
%YTf&rA@c(?x)4GV% A?!45?!%YTf&'2??
TQ? Actively cooled
jacket according to
?0QX`S??C4ri ?@Y?X0?0IB#W??6r?P1u4P`"Qp ?F?#W???0IB
TQ? Evaporation unit
actively controls
continuous flow
of liquid nitrogen: most stable
temperature signal for low-temperature applications
Build on Your Rheometer:
Application-specific Accessories
Structure Analysis
Gather sample structure information
by combining these optical and
dielectric methods with rheology.
Extended Material
Characterization
These accessories transfer the
MCR rheometer’s measuring
capabilities into other material
characterization applications.Obtain additional structure information, set additional parameters or employ the rheometer’s
functionality for further material characterization: This wide range of application-specific
accessories is easily integrated into your MCR rheometer.
Rheo-Microscopy
(Fluorescence,
Polarized,
Non-Polarized)
Small-angle light
scattering (SALS)
Small-angle X-ray
scattering (SAXS)Small-angle
neutron
scattering
(SANS) Particle image
velocimetry (PIV)
Polarized Imaging
Dielectro-
Rheological Device
(DRD)
Pressure cells UV Curing SystemImmobilization CellMagneto-
Rheological
Device Electro-Rheological
Device
Humidity Option
for CTD 180
Extensional
rheology Dynamic
mechanical thermal
analysis (DMTA)Starch rheology
Large-particle
rheologyInterfacial rheology
Tribology:Ball on three plates
Pin on disk
Four ball Powder Cell
Additional
Parameter Setting
Employ these accessories to set
additional parameters together with
the temperature for rheological
tests.
Extended Material
Characterization
MCR
MCR
Application-
specific
Accessories
for Additional
Parameter
Setting
MCR
Application-specific
Accessories for
Structure Analysis
15
14
The Advantage of Diversity
MCR Measuring SystemsMeasuring System Variants
The MCR measuring systems can be used with all
temperature devices and are interchangeable within
their category of accessories. For example, a PP25 parallel-plate measuring system can be used in all
according LTD, PTD, ETD or CTD systems. All geometry dimensions, safety limitations and
calibration constants are saved in the Toolmaster™
chip located in the coupling of every measuring system.
Made from diverse materials and featuring different
surfaces and dimensions, all measuring systems are optimized regarding compliance, thermal expansion and thermal conductivity. With hundreds and hundreds of measuring systems,
and their efficient combination with a wide range of
environmental systems, there is barely any application
that cannot be covered by an Anton Paar MCR rheometer. The following nomenclature gives you an overview of
available measuring system variants and how they are
denoted.
Concentric
!3Uy? b Double Gap
?&q4Cone-Plate
Wrw= Parallel Plate
' fI? Tack
SystemsTribology
Systems Stirrers
Low-
viscosity liquids Viscoelastic
liquids Melts
Paste-like
materials Gel-like
materials Soft solids
!Sf?
reactive systems Solids
DMTA
Torsion DMTA
Extension
D !bbRD P P !bbRD !bbRD P R / S S / S
Dimension in mm
Prefix
D
::: Disposable plate
DC ::: Disposable cup
FDD ::: Fixture for disp. dishes
DD ::: Disposable dish
CAP ::: Cap plate
Type
CC ::: Concentric cylinder
CPP ::: Cone Partitioned Plate
DG ::: Double gap
ST ::: Stirrer
PP ::: Parallel plate
CP ::: Cone-plate
ME ::: Mooney Ewart
BM ::: Ball measuring system
PPR ::: Plate-plate ring
CPR ::: Cone-plate ring SRF
::: Solid rectangular fixture
SCF ::: Solid circular fixture
UXF ::: Universal extensional fixture
SER ::: Sentmanat ext. rheometer
BIC ::: Bi-cone
BC ::: Tribology system
TG ::: Twin gap
MD ::: Measuring dish
Shaft
PR ::: Pressure
Z ::: Zerodur shaft
PE ::: PEEK shaft
ERD ::: Electro-Rheological Device
MRD ::: Magneto-Rheological Device
TG ::: TruGap™
DI ::: Dielectro-Rheological Device
CTD ::: Convection Temperature Device
Surface
S ::: Sandblasted
P2% A? qYDb% A? 5#QT?F("6? "GB@7@?IP#Apyb
P3% A? qYDb% A? 5#QT?F(H??`"GB@77??09??R7
P6% A? qYDb% A? 5#QT?F(&re)'bU!#!S?'3?#)???@?IP9??RT
P7% A? qYDb% A? 5#QT?F(XUDg'bU!#!Sf()#)???@?IP#Apyb
PX ::: Profile special
HL ::: Helical profile left handed
HR ::: Helical profile right handed
HX ::: Helical profile special
CX ::: Coated
Material
SS ::: Stainless steel
TI ::: Titanium
HA ::: Hastelloy
INV ::: Invar
PC ::: Polycarbonate
AL ::: Aluminum
GL ::: Glass
INC ::: Inconel
CA ::: Carbon
17
16
UnitM C R 102 MCR 302MCR 502 TDR
Bearing -Air Air Air
EC motor (brushless DC@?…P with high-resolution optical encoder -
Permanent torque (60 min@?…P, no signal drift-
EC mode (controlled shear rate and shear stress@?…P
Digital Eye, Software video option and camera-
CoolPeltier'BX?, Peltier system with built-in cooling option requiring no
additional accessories for counter-cooling ??(0
-5 to 200 -5 to 200 -5 to 200
Peltier-based convection oven, does not require LN
2 for cooling ??(0-20 to 180 -20 to 180 -20 to 180
Actively Peltier-controlled hood, Peltier technology ??(06r?P#F??R5?wr@p?0@X ?F?#W??#W?? 6r?P#F??R5?wr@p?0@X ?F?#W??#W?? 6r?P#F??R5?wr@p?0@X ?F?#W??#W??
Concentric-cylinder Peltier temperature control ??(0-30 to 200 -30 to 200 -30 to 200
Maximum temperature range ??(0-160 to 1000 -160 to 1000 -160 to 1000
Pressure range barup to 1000 up to 1000 up to 1000
?? TP!45?!@p?&?@?C0BGrA@c#?iy"% A?AG)?rA@c94h)% A?q1`dAp?H13?&?@?C5?UB BRG%YTf&GV@p?$?v?0@?% A??? TP$R?6B BRG?? TPC XB`4p?? -
TruGap'BX? for in-place measurement and control of the gap -
TwinDrive-ready -
&?T5'Y %`? GV!t "C@p?% A?%YTf&?H13?rA@c'Y `dApB BRG%YTf&?H13?GV%YTf&%YTf&% A?rA@c0BGPw"?i)w !45?!!?ui% A?q1`dAp?H13??7T PT6%`? -
Tr u R a t e'BX? -
TruStrain'BX? -
Normal force and velocity profiles, tack, squeeze -
Raw data (L AOS, waveform, ...@?…P-
With Exposed Suppor t Plate (WESP@?…P-
optional
Pushing the Boundaries of Rheometry:
MCR Rheometers Achieve Lowest Ever Torque
Enhanced low-torque performance in
shear-rate-controlled tests
A shear-rate-controlled rotational test in the figure below
shows the certified viscosity of a standard oil within 5 %
down to a torque of 1 nNm. This screenshot is taken directly
from the software; every single point is displayed, showing
the equidistance of point distribution in a precise shear-rate-
controlled test without any extrapolation.
Specifications
TruStrain TM control with increased data accuracy
The figure below shows a strain sweep in strain control down
%$Q?8 ?02u"?i?8cda?5?G?$D?4r#xSw?h6GF't?FR1?Ei?4ff??@t? tWFW
also an actual measurement without any further processing or
hidden data points. Therefore the data points are equidistantly
distributed.
Measurements tell you more than words
The new specifications of Anton Paar’s rheometers are shown
by a single measurement. Come to one of Anton Paar’s various
demonstration labs worldwide – let experts measure your sample
and discuss your specific application. Anton Paar is ready to help
you fulfill your application requirements.
MCR rheometers are constantly being improved by a dedicated development team
continuously furthering the rheometers’ core components. Anton Paar now reports a
significant breakthrough in the minimum torque levels that can be measured – see data below.
The technology developed for the groundbreaking TwinDrive system has been applied to
the controller at the core of every MCR motor, and new production processes have been
introduced. This means that the entire MCR series can now provide results of even greater
precision, which is reflected in a new set of specifications.
19
18
www.anton-paar.com
AVH'9 2020 Anton Paar GmbH | All rights reserved.
Specifications subject to change without notice.
C92IP001EN-M
Entreprise(s) concernée(s) :
Produit(s) concerné(s) :
Date d'upload du document :
vendredi 8 janvier 2021