Document
H54IP001EN-E_Datasheet_BPA5
Contenu du document
Consistency & Ductility
BPA 5
Fraass Breaking Point Tester
H54IP001EN-E
BPA 5
Fraass Breaking Point Tester
The automatic breaking point tester determines the brittle
behavior of bitumen at low temperatures.
The Fraass breaking point is the temperature at which
the first crack appears in the coating of a thin, flat steel
plaque, flexed under descending temperatures.
Benefits at a Glance
- Peltier cooling for the refrigeration of the test
chamber
- Touch-key panel with large LC-display
- Software BPACon (optional)
- Calibration set (optional)
- Melting apparatus BPM 5 (optional) for steel plaque
coating
Convenient Operation
Test chamber refrigeration by Peltier elements. These
solid-state cooling elements require a light auxiliary
refrigerator only and you therefore avoid investing in a
bulky and highly energy-consuming cooler.
Customized User Flexibility
In connection with the BPACon software the breaking
point tester serves as a research tool for different
coatings. Not only test proceedings (e.g. variation of
the bending force) can be easily watched via a graphic
diagram on the PC screen, it is also possible to run
Permanent Bending Tests in which the trend of the
bending force vs. the temperature drops.
The Force Recovery Trend serves as the criteria for the
elasticity of the material under test. The progression in
plasticity and in cracking will be displayed numerically
or graphically in different scales. With specially created
curves of trends for bending or recovery force even very
small, nearly not visible breaks of the material can be
found with high precision. Standard Methods
EN 12593, JIS K 2207, IP 80
Technical Specifications
Application range -45 °C to 60 °C (dependent on the
temperature of the cooling circuit)
Programs
- Standard (cooling rate 1 K/min)
for Fraass breaking point
- Rapid (cooling rate 2 K/min)
for very low breaking points
- Search mode
to find the approx. breaking point
- User-definable
to set up program modifications
Refrigeration Peltier elements and optional circulation
cooler
Detection principle Force (resolution 10 mN)
Interfaces RS232 for printer, RS232 for PC,
LIMS compatibility
Power supply 115 V/230 V, 50 Hz/60 Hz, 450 W
Dimensions 230 mm x 450 mm x 400 mm (W x D x H)
Weight approx. 12.5 kg
Resolution 0.1 °C
? 2019 Anton Paar GmbH | All rights reserved.
Specifications subject to change without notice.
Your distributor:
Entreprise(s) concernée(s) :
Produit(s) concerné(s) :
Date d'upload du document :
mercredi 6 janvier 2021